Diagnostic X-ray tube assembly IRD-67

The x-ray tube assembly is designed for integration in the production of new medical radiographic and fluoroscopic stationary and mobile devices, as well as for installation in compatible devices in case of replacement of the transmitter during repair and maintenance of the majority of X-ray devices in service.

The X-ray tube assembly is equipped with an X-ray tube manufactured by “Svetlana- X-ray” JSC with focal spots of 0.6 mm and 1.2 mm at an allowable maximum voltage of 150 kV. The tube type is 27-75BD67-150. X-ray tube assembly connection is compatible with IEC 60526 high-voltage cable connectors.

Table 1: Main technical specifications

DescriptionMeasurement
Nominal X-ray tube voltage:150 kV
Nominal focal spot size 
– large focus1,2 mm
– small focus0,6 mm
Nominal anode input power (at 0,1 sec), 50Hz: 
– large focus40 kW
– small focus14 kW
Nominal anode input power (at 0,1 sec), 60Hz: 
– large focus44,6 kW
– small focus16 kW
Nominal anode input power (at 0,1 sec), 180Hz: 
– large focus75 kW
– small focus27 kW
  

 

Table 2: Maximum and minimum parameter values.

DescriptionMeasurement
Nominal X-ray tube voltage 
– Radiographic150 kV
– between anode (or cathode) and ground75 kV
Minimum X-ray tube voltage40 kV
Maximum X-ray tube Current 
– large focus1000 mA
– small focus400 mA
Maximum filament current : 
– large focus5,5 А
– small focus5,5 А
Filament voltage 
– large focus (at max. filament current 5,5A)15,9 V
– small focus (at max. filament current 5,5A)9,0 V 
Filament frequency limit0-25 kHz
Continuous anode input power200 W (283 HU/s)
Thermal characteristics: 
– anode heat content 210 kJ (300 kHU)
– maximum anode heat dissipation500 W
– X-ray tube assembly heat content900 kJ (1250 kHU)
Nominal continuous input powe 
– without air-cooling180 W (15 kHU/min)
Operating limits 
– temperature10-40 °С

 

Table 3: Stator specifications

DescriptionMeasurement
OverclockingUnderclocking
Rotational speed (Hz)18050/6018050/60
Input power (W)2300145030080
Voltage (V)40024013058
Current (А)5,46,521,5
Minimum acceleration time (sec)10,6
Capacity (µF)324324

 

The IRD-67 X-ray tube assembly of all versions can be equipped with a stator with the following parameters:

Table 4: Stator specifications

DescriptionMeasurement
Overclocking Underclocking
Rotational speed (Hz)18050/6018050/60
Input power (W)11009108383
Voltage (V)2201306040
Current (А)5,77,81,62,3
Minimum acceleration time (sec)1,20,8
Capacity (µF)644644

 

Table 5: Mechanical specifications

DescriptionMeasurement
Anode: 
Angle12 °
Diameter72 mm
Materialtungsten-rhenium-molybdenum
Radiation filtering: 
– permanent filtration1 mm Al / 75 kV IEC 60522
High voltage connectorsaccording to IEC 60526-1:2010
Refrigerationnatural or forced air